Practical Sterilization Function:
1. Preset Vacuum Function is able to shorten the sterilization time.
2. Vacuum drying function is able to remove unwanted moisture quickly, and to reach the Bowie Dick test criteria.
3. Cloth / apparatus & glass / liquid, Sterilization Functional Switch Mode can be selected to switch according to different kinds of materials to be sterilized. When liquid sterilization, make sure liquid ingredient to avoid rust and corrosion occurred. For medicalt & chemical ingredient, physiological salt water etc., please select stainless steel #316.
4. Auto Water Supply Device: Electrical control system is able to detect and sense the water volume automatically. Upon insufficient water volume, it will supply to high water level automatically. Use equipment for soft water purification can prolong the life cycle of sterilizer.
Safety Devices:
1. Auto Power-Off Protection Device for water shortage, when water runs out, the electrical control system will automatically cut off the heating power to prevent heating device from overheating, dry-out or even meltdown.
2. Pressure Controller, the maximum pressure in the boiler is limited to 2.2 ±0.1kg/cm 2. When the pressure reaches the set value, the heating power will be automatically off to prevent the pressure from rising.
3. Safety Valve, when the pressure reaches 2.5kg/cm 2, it will start a rapid exhaust of steam and release the pressure from the chamber .
4. Pressure Lock, when the pressure of inner chamber exceeds 0.2kg/cm2, the clutch will be locked automatically, and the door cover will not be able to open, in order to keep the operator safe.
5. Safety Switch of Door Cover, when the door cover opened, the steam in the outer chamber is unable to enter the inner chamber. It cannot be operated normally unless the door cover is closed tightly.
|